        <article-title>Repeatability of Noninvasive Break-Up Time Measurements using Keratograph Oculus 3</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <name>
            <surname>Musilová</surname>
            <given-names>Lenka</given-names>
          </name>
          <contrib-id contrib-id-type="orcid">https://orcid.org/0000-0003-1305-6153</contrib-id>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Vodičková</surname>
            <given-names>Alena</given-names>
          </name>
        </contrib>
        <contrib contrib-type="author">
          <name>
            <surname>Pluháček</surname>
            <given-names>František</given-names>
          </name>
          <contrib-id contrib-id-type="orcid">https://orcid.org/0000-0001-7551-0070</contrib-id>
        </contrib>
      </contrib-group>
      <pub-date date-type="pub" publication-format="electronic">
        <day>6</day>
        <month>6</month>
        <year>2024</year>
      </pub-date>
      <issue>5</issue>
      <elocation-id>5</elocation-id>
      <abstract>
        <p>Aim: The primary aim of this study is to evaluate the repeatability of noninvasive break-up time (NIBUT) measurement by keratograph when it is determined from one, two or three partial measurements, and to recommend a suitable methodology for practice. Another goal is to verify that repeated measurements do not affect the measured value. Material and Methods: Thirty-eight healthy volunteers (30 women and 8 men) aged between 19 and 50 years old were included in the study, in which only one eye of each volunteer was measured. The study was designed as a prospective one. Each subject adapted to the local conditions of the laboratory for 15 minutes and subsequently underwent two series of NIBUT measurements (test, retest) on an OCULUS 3 Keratograph. The minimum time interval between the two series was 10 minutes, in which each series contained three partial measurements approximately 3 three measurements in the given series. Repeatability was assessed by a Bland-Altman analysis and expressed as a repeatability coefficient. In every case, only the time of the first break-up of the tear film was monitored. Results: The statistical analysis did not show statistically significant differences both between partial measurements of NIBUT in the individual series (p = 0.92, p = 0.81) and when comparing all six measurements (p = 0.95). The mean values of the partial measurements ranged from 13.6 s to 14.4 s. The repeatability coefficients were found to be 15.0 s, 12.1 s and 10.0 s for methodologies A, B and C, respectively. A supplementary analysis for 12 eyes with low NIBUT (&lt; 10 s) showed statistically significantly better repeatability in this group, with coefficients of 7.0 s (methodology A), 6.0 s (B) and 4.6 s (C). Conclusion: Determination of NIBUT from three consecutive measurements (with a sufficient interval of ideally a few minutes) significantly improves repeatability. Such repeated NIBUT measurements do not have a significant effect on the measured value. The mentioned methodology for measuring NIBUT on a keratograph can be recommended for use in practice.</p>
      </abstract>
      <kwd-group>
        <kwd>break-up time test</kwd>
        <kwd>BUT</kwd>
        <kwd>NIBUT</kwd>
        <kwd>repeatability of measurement</kwd>
        <kwd>tear film</kwd>
        <kwd>dry eye syndrome</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <back>
    <ref-list>
      <title>References</title>
      <ref id="R2706">
        <mixed-citation>Clayton JA. Dry eye. N Engl J Med. 2018;378:2212-2223. doi: 10.1056/NEJMra1407936<pub-id pub-id-type="doi">10.1056/NEJMra1407936</pub-id></mixed-citation>
      </ref>
      <ref id="R2707">
        <mixed-citation>Uchino M, Schaumberg DA. Dry eye disease: impact on quality of life and vision. Curr Ophthalmol Rep. 2013;1:51-11. doi: 10.1007/ s40135-013-0009-1</mixed-citation>
      </ref>
      <ref id="R2708">
        <mixed-citation>Craig JP, Nelson JD, Azar DT et al. TFOS DEWS II report executive summary. Ocul Surf. 2017;15:802-812. doi: 1016/j.jtos.2017.08.003</mixed-citation>
      </ref>
      <ref id="R2709">
        <mixed-citation>Messmer EM. The Pathophysiology, Diagnosis, and Treatment of Dry Eye Disease. Dtsch Arztebl Int. 2015;112:71-82. doi: 10.3238/ arztebl.2015.0071</mixed-citation>
      </ref>
      <ref id="R2710">
        <mixed-citation>Wolffsohn JS, Arita R, Chalmers R et al. TFOS DEWS II diagnostic methodology report. Ocul Surf. 2017;15:539-574. doi: 10.1016/j. jtos.2017.05.001<pub-id pub-id-type="doi">10.1016/j</pub-id></mixed-citation>
      </ref>
      <ref id="R2711">
        <mixed-citation>Mengher LS, Bron AJ, Tonge SR, Gilbert DJ. Effect of fluorescein instillation on the pre-corneal tear film stability. Curr Eye Res. 1985;4:9-12. doi: 10.3109/02713688508999961<pub-id pub-id-type="doi">10.3109/02713688508999961</pub-id></mixed-citation>
      </ref>
      <ref id="R2712">
        <mixed-citation>Patel S, Murray D, McKenzie A, Shearer DS, McGrath BD. Effects of fluorescein on tear breakup time and on tear thinning time. Am J Optom Physiol Opt. 1985;62:188-190. doi: 10.1097/00006324- 198503000-00006<pub-id pub-id-type="doi">10.1097/00006324-</pub-id></mixed-citation>
      </ref>
      <ref id="R2713">
        <mixed-citation>Pult H, Riede-Pult BH. A new modified fluorescein strip: its repeatability and usefulness in tear film break-up time analysis. Cont Lens Anterior Eye. 2012;35:35-38. doi: 10.1016/j.clae.2011.07.005<pub-id pub-id-type="doi">10.1016/j.clae.2011.07.005</pub-id></mixed-citation>
      </ref>
      <ref id="R2714">
        <mixed-citation>DEWS Diagnostic Methodology, 2007a. The definition and classification of dry eye disease report of the Definition and Classification Subcomittee of the International Dry Eye Workshop.</mixed-citation>
      </ref>
      <ref id="R2715">
        <mixed-citation>Finis D, Ackermann P, Pischel N et al. Evaluation of meibomian gland dysfunction and local distribution of meibomian gland atrophy by non-contact infrared meibography. Curr Eye Res. 2015; 40:982-989. doi: 10.3109/02713683.2014.971929<pub-id pub-id-type="doi">10.3109/02713683.2014.971929</pub-id></mixed-citation>
      </ref>
      <ref id="R2716">
        <mixed-citation>Korb DR. The Tear Film: structure, function and clinical examination. 1st ed. Oxford: Butterworth-Heinemann/BCLA; 2002. 208. ISBN 0-7506-4196-7</mixed-citation>
      </ref>
      <ref id="R2717">
        <mixed-citation>Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8:307-310. doi: 10.1016/S0140-6736(86)90837-8<pub-id pub-id-type="doi">10.1016/S0140-6736(86)90837-8</pub-id></mixed-citation>
      </ref>
      <ref id="R2718">
        <mixed-citation>Carkeet A. Exact parametric confidence intervals for Bland-Altman limits of agreement. Optom Vis Sci. 2015;92:e71-80. doi: 10.1097/ OPX.0000000000000513 14 Cho P, Leung L, Lam A, Choi A. Tear break-up time: clinical procedures and their effects. Ophthalmic Physiol Opt. 1998;18:319-324. doi: 10.1046/j.1475-1313.1998.00385.x<pub-id pub-id-type="doi">10.1046/j.1475-1313.1998.00385.x</pub-id></mixed-citation>
      </ref>
      <ref id="R2719">
        <mixed-citation>Nichols KK, Mitchell GL, Zadnik K. The repeatability of clinical measurement of dry eye. Cornea. 2004; 23:272-285. doi: 10.1097/00003226-200404000-00010<pub-id pub-id-type="doi">10.1097/00003226-200404000-00010</pub-id></mixed-citation>
      </ref>
      <ref id="R2720">
        <mixed-citation>Bandlitz S, Peter B, Pflugi T et al. Agreement and repeatability of four different devices to measure non-invasive tear breakup time (NIBUT). Cont Lens Anterior Eye. 2020; 43:507-511. doi: 10.1016/j. clae.2020.02.018<pub-id pub-id-type="doi">10.1016/j</pub-id></mixed-citation>
      </ref>
      <ref id="R2721">
        <mixed-citation>Fernández J, Rodríguez-Vallejo M, Martínez J et al. Greafe’s Arch Clin Exp Ophthalmol 2018. Agreement and repeatability of objective systems for assessment of the tear film. Graefes Arch Clin Exp Ophthalmol. 2018; 256:1535-1541. doi: 10.1007/s00417-018-3986-9<pub-id pub-id-type="doi">10.1007/s00417-018-3986-9</pub-id></mixed-citation>
      </ref>
      <ref id="R2722">
        <mixed-citation>Lemp MA, Hamill JR. Factors affecting tear film breakup in normal eyes. Arch Ophthalmol. 1973; 89:103-105. doi: 10.1001/archopht.1973.01000040105007<pub-id pub-id-type="doi">10.1001/archopht.1973.01000040105007</pub-id></mixed-citation>
      </ref>
      <ref id="R2723">
        <mixed-citation>Tian L, Qu JH, Zhang XY, Sun XG. Repeatability and Reproducibility of Noninvasive Keratograph 5M Measurements in Patients with Dry Eye Disease. J Ophthalmol. 2016; 2016:8013621. doi: 10.1155/2016/8013621<pub-id pub-id-type="doi">10.1155/2016/8013621</pub-id></mixed-citation>
      </ref>
      <ref id="R2724">
        <mixed-citation>García-Marqués JV, Martínez-Albert N, Talens-Estarelles C et al. Repeatability of Non-invasive Keratograph Break-Up Time measurements obtained using Oculus Keratograph 5M. Int Ophthalmol. 2021; 41:2473-2483. doi: 10.1007/s10792-021-01802-4<pub-id pub-id-type="doi">10.1007/s10792-021-01802-4</pub-id></mixed-citation>
      </ref>
      <ref id="R2725">
        <mixed-citation>Singh S, Srivastas S, Modiwala Z et al. Repeatability, reproducibility and agreement between three different diagnostic imaging platforms for tear film evaluation of normal and dry eye disease. Eye. 2023; 37:2042-2047. doi: 10.1038/s41433-022- 02281-2<pub-id pub-id-type="doi">10.1038/s41433-022-</pub-id></mixed-citation>
      </ref>
      <ref id="R2726">
        <mixed-citation>Nichols JJ, Nichols KK, Puent B, Saracino M, Mitchell GL. Evaluation of tear film interference patterns and measures of tear breakup time. Optom Vis Sci. 2002;79:363-369. doi: 10.1097/00006324- 200206000-00009<pub-id pub-id-type="doi">10.1097/00006324-</pub-id></mixed-citation>
      </ref>
      <ref id="R2727">
        <mixed-citation>Hong J, Sun X, Wei A et al. Assessment of tear film stability in dry eye with a newly developed keratograph. Cornea. 2013;32:716-</mixed-citation>
      </ref>
      <ref id="R2728">
        <mixed-citation>Lan W, Lin L, Yang X, Yu M. Automatic noninvasive tear breakup time (TBUT) and conventional fluorescent TBUT. Optom Vis Sci. 2014; 91:1412-1418. doi: 10.1097/OPX.0000000000000418<pub-id pub-id-type="doi">10.1097/OPX.0000000000000418</pub-id></mixed-citation>
      </ref>
      <ref id="R2729">
        <mixed-citation>Itokawa T, Suzuki T, Koh S, Hori Y. Evaluating the Differences Between Fluorescein Tear Break-up Time and Noninvasive Measurement Techniques. Eye Contact Lens. 2023; 49: 104-109. doi: 10.1097/ICL.0000000000000966<pub-id pub-id-type="doi">10.1097/ICL.0000000000000966</pub-id></mixed-citation>
      </ref>
      <ref id="R2730">
        <mixed-citation>doi: 10.1097/ICO.0b013e3182714425<pub-id pub-id-type="doi">10.1097/ICO.0b013e3182714425</pub-id></mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>